Huis > producten > Digitale Stroommeter >
Industrial-Grade Coriolis Flowmeter Promass 8I1B Portable Coriolis Meter From Germany's Endress Hauser-Customizable OEM Support

Industrial-Grade Coriolis Flowmeter Promass 8I1B Portable Coriolis Meter From Germany's Endress Hauser-Customizable OEM Support

Industrial-Grade Coriolis flowmeter with OEM support

Portable Coriolis meter from Endress Hauser

Digital flow meter with customizable options

Plaats van herkomst:

CHINA

Merknaam:

E+H

Modelnummer:

8i1b

Praatje Nu
Vraag een offerte
Productgegevens
Voedingsspanning:
10-30 VDC
Het meten van principe:
Absolute en meter druk
Procesverbinding:
1/2 "NPT
Montagetype:
Schroefdrend
Elektrische verbinding:
M12x1
Referentie -nauwkeurigheid:
Standaard: 0,05% platina: tot 0,025%
Uitgangssignaal:
4-20 ma
Montagemethode:
DIN -rail of wandbevestiging
Protocol:
Hart/RS485
Temperatuurbereik:
-40 tot 185°F.
Bereik:
0-10.000 psi
Bevochtigde onderdelen materiaal:
Roestvrij staal
Materiaal:
Roestvrij staal
Communicatieprotocol:
In een deksel
Bedrijfstemperatuur:
-20 tot 85 graden Celsius.
Markeren:

Industrial-Grade Coriolis flowmeter with OEM support

,

Portable Coriolis meter from Endress Hauser

,

Digital flow meter with customizable options

Betalings- en verzendvoorwaarden
Min. bestelaantal
1 pc
Prijs
USD 600
Levertijd
7-14 dagen
Betalingscondities
T/t
Levering vermogen
100 pc
Productbeschrijving

What is the Promass I 100

The Promass I 100 is a high-performance Coriolis mass flowmeter designed for industrial applications, offering measurements of mass flow, density, and temperature, with optional viscosity measurement.

It is part of the Endress+Hauser “Promass I” line, featuring a straight single-tube sensor design and very compact transmitter housing.

 

Measuring principle

  • The fundamental principle is Coriolis force measurement: the flow tube is vibrated and the fluid flowing through causes a phase shift in oscillations between inlet & outlet positions. This phase shift is proportional to mass flow.
  • The tube is excited at its resonance; changes in oscillation frequency due to changes in mass (tube + fluid) are used to compute density.
  • Temperature of the measuring tube is also measured, both for fluid temperature and to compensate temperature effects.

 

Key Technical Features & Specifications

Here are some of its important characteristics:

Feature

Details

Flow variables measured

Mass flow, density, temperature; optional viscosity.

Sensor design

Straight single-tube design; “easy cleanable”.

Tube material / wetted parts

Titanium for measuring tube; other parts in stainless steel or coating depending on version.

Size range (nominal diameter)

DN 8 to DN 80 (≈ 3/8″ to 3″).

Accuracy

• Liquids: about ±0.1 % for mass & volume flow; <-- density accuracy in liquid ~ ±0.0005 g/cm³.
• Gases: larger error, e.g. approx ±0.5 % mass flow.

Temperature / Pressure ranges

Medium temperature: from 50 °C up to approx +150 °C.
Process pressure: up to PN100 (or equivalent), specifics depending on DN, material etc.

Environmental / Protection

High ingress protection (IP69K in many versions). Robust transmitter housing.

Installation

“Full-bore” design to minimize pressure drop. As its straight tube, less demanding for inlet/outlet runs compared to bent-tube designs.

Outputs / Communication

Multiple typical industrial outputs: analog (4-20 mA with HART), digital protocols (Modbus RS-485, PROFIBUS, PROFINET, etc.), pulse/frequency/switch outputs.

Additional features

Integrated web server for local configuration, diagnostics. “Heartbeat Technology” for verification and monitoring.

 

Advantages

  • High accuracy and stability: Since the measurement is mass‐based and not strongly influenced by fluid viscosity, density, or flow profile, it gives very reliable readings.
  • Multi-variable measurement: Flow, density, temperature (and viscosity optionally) reduce need for multiple sensors.
  • Compact & robust: Small footprint transmitter, full-bore, straight tube helps in installations where space is limited and/or cleaning is needed.
  • Low pressure loss: Full bore and straight single tube help minimizing pressure drop. Good for sensitive fluids.
  • Good for hygiene and cleaning: Because of the easy-clean straight tube. Also options exist for hygienic/stainless housing.

 

Limitations / Considerations

  • Cost tends to be relatively high vs simpler flowmeters. More sophisticated construction, more electronics.
  • For very low flow or very small diameters, sensitivity and signal-to-noise becomes more challenging need to check minimum flow (cut-off) for your application.
  • The wetted materials (Titanium, stainless) have to be compatible with the fluid (chemically, abrasive, etc.). If the fluid is aggressive or contains solids, consider wear or corrosion.
  • Temperature and pressure limits: depending on the exact model; ensure your process operating envelope fits within them.
  • Maintenance might require more care (cleaning, calibration) due to the precision of the instrument.

 

Typical Applications

The Promass I 100 is suitable for many industrial sectors, especially where accurate mass flow + density + temperature are needed, such as:

  • Chemical process lines
  • Food & beverage where hygiene / cleaning is required
  • Pharmaceuticals
  • Oil & gas (for certain liquids/gases)
  • Skid / module builders (because of compact size)
  • Any application where pressure drop must be minimized or full-bore is preferable.

 